Comparison of Central Corneal Thickness and Anterior Chamber Depth Measured Using Three Different Devices.
Se Hun Han, Ho Sik Hwang, Min Chul Shin, Kyung Eun Han
Department of Ophthalmology, Chuncheon Sacred Heart Hospital, Hallym University College of Medicine, Chuncheon, Korea. hkeoph@gmail.com
서로 다른 세 가지 기기로 측정한 중심각막두께 및 전방깊이의 비교
한세훈⋅황호식⋅신민철⋅한경은
한림대학교 의과대학 춘천성심병원 안과학교실
Abstract
PURPOSE
To compare measurements of central corneal thickness (CCT) and anterior chamber depth (ACD) obtained using Galilei(TM), Pentacam(R) (Oculus, Wetzlar, Germany) and Lenstar(R) (Haag-Streit, Koeniz, Switzerland) and analyze the measurement agreements. METHODS: CCT and ACD were measured using Galilei(TM), Pentacam(R) and Lenstar(R) in 47 eyes of 25 healthy subjects. The measurements were compared among the 3 devices. RESULTS: The average CCT measurements using Galilei(TM), Pentacam(R) and Lenstar(R) were 552.6 +/- 29.41 microm, 543.9 +/- 30.50 microm and 537.5 +/- 30.26 microm, respectively. The measurements significantly correlated with each other (r > 0.9, p < 0.001), but were statistically significantly different (p < 0.001). The average ACD measurements using Galilei(TM), Pentacam(R) and Lenstar(R) were 3.23 +/- 0.360 mm, 3.22 +/- 0.403 mm and 3.19 +/- 0.367 mm, respectively. The measurements significantly correlated with each other (r > 0.9, p < 0.001), but were statistically significantly different (p = 0.034). The CCT 95% limits of agreement (LoA) between Galilei(TM) and Pentacam(R), Pentacam(R) and Lenstar(R) and Lenstar(R) and Galilei(TM) were 31.95 microm, 44.76 microm and 46.57 microm, respectively and 95% ACD LoA were 0.46 mm, 0.32 mm and 0.28 mm, respectively. CONCLUSIONS: CCT and ACD measured using the 3 devices were highly correlated with each other but the measurements were statistically different. Therefore, the measurements were not interchangeable and these differences should be considered in clinical use.
Key Words: Anterior chamber depth;Central corneal thickness;Galilei(TM);Lenstar(R);Pentacam(R)


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next