Journal of the Korean Ophthalmological Society 2007;48(6):860-865.
Published online June 30, 2007.
A Wide Discrepancy between Intraocular Pressure by Applanation and Non-contact Tonometry after LASIK.
Jin Chul Kim, Tae Young Chung, Chang Won Kee
Department of Ophthalmology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea. cwkee@smc.samsung.co.kr
테논낭의 회전 피판을 이용한 저안압 치험 1예
김연희,김상국,형성민
Department of Ophthalmology, School of Medicine, Chungbuk National University, Chungbuk, Korea
Correspondence:  Younhui Kim, M.D.
Abstract
PURPOSE
To report the case of a glaucoma patient with a history of LASIK surgery who presented with a wide discrepancy in intraocular pressure between the Goldmann applanation tonometry and non-contact tonometry. METHODS: A 20-year-old male glaucoma patient was referred to our clinic with a chief complaint of blurred vision in his right eye. The patient had a history of uneventful bilateral LASIK surgery. Upon our initial examination, his best-corrected visual acuity in the right eye was 0.04 and his right eye IOP was 14 mmHg by Goldmann applanation tonometry and 40 mmHg by non-contact tonometry. A slit lamp examination showed corneal epithelial edema with diffuse interface opacities. The central corneal thickness in the right eye was 587 micrometer. An ultrabiomicroscopy demonstrated an interface fluid accumulation between the LASIK flap and the stromal bed. The patient was diagnosed with juvenile open-angle glaucoma and was scheduled for a trabeculectomy. RESULTS: Seven days after trabeculectomy, the patient's vision in the right eye improved to 0.7 and his IOP was 12 mmHg by Goldmann applanation tonometry and 14 mmHg by non-contact tonometry. In addition, the corneal epithelial edema, interface opacities, and fluids had all resolved and the central corneal thickness was reduced to 489 micrometer. CONCLUSIONS: In eyes that have undergone LASIK surgery, the IOP measured by Goldmann applanation tonometry may underestimate the true pressure due to fluid accumulation under the LASIK flap. In such cases, non-contact tonometry can be a convenient and a reliable method for determining treatment plan.
Key Words: Goldmann applanation tonometry;Interface fluid;LASIK;Non-contact tonometry


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next