Clinical Evaluation of A Noncontact Applanation Tonometer.
Chul Hong
Department of Ophthalmology, School of Medicine, Hanyang University, Seoul, Korea.
비접촉성안압계를 이용한 안압측정
홍철 ( Chul Hong )
Abstract
A new applanation tonometer(American Optical Corp.) measuring intraocular pressure without contact between the instrument and the eye was developed recently. Applanation is produced by a controlled air-pulse of linearly increasing force impinging on the cornea. To observe clinical evaluation of this noncontact tonometer, the intraocular pressures of 199 eyes of one hundred and one men and women at the age of 14 to 60 year without corneal opacity and corneal edema were measured with the Goldmann applanation tonometer(control group), the noncontact tonometer and the Schitotz tonometer and the correlations between the different instruments were investigated with comparison of the measured values. The results were as follows: 1. Comparing the mean values, the noncontact tonometer was more similar to the Goldmann 'applanation tonometer than the Schitotz tonometer. 2. Comparing the correlation coefficients and the differences between pairs of measuring instruments, the noncontact tonometer was more highly correlated to the Goldmann applanation tonometer than the Schitotz tonometer. Therefore, the noncontact tonometer can be highly evaluated for clinical applications-especially frequent measurements at short intervals and glaucoma screening programs-due to its accuracy, simplicity, painlessness and repeatability.


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next