J Korean Ophthalmol Soc > Volume 57(3); 2016 > Article
Journal of the Korean Ophthalmological Society 2016;57(3):369-379.
DOI: https://doi.org/10.3341/jkos.2016.57.3.369    Published online March 15, 2016.
Intraocular Lens Power Calculations Using Dual Scheimpflug Analyzer.
Young Bok Lee, Ka Young Yi, So Hyun Bae, Ha Kyoung Kim, Joon Young Hyon, Won Ryang Wee, Young Joo Shin
1Department of Ophthalmology, Kangnam Sacred Heart Hospital, Hallym University College of Medicine, Seoul, Korea. schinn@hanmail.net
2Department of Ophthalmology, Seoul National University College of Medicine, Seoul, Korea.
3Department of Ophthalmology, Seoul National University Bundang Hospital, Seoul National University College of Medicine, Seongnam, Korea.
이중샤임플러그 전안부분석기를 이용한 백내장 수술 시 인공수정체 도수 계산
이영복1⋅이가영1⋅배소현1⋅김하경1⋅현준영2,3⋅위원량2⋅신영주1
한림대학교 의과대학 강남성심병원 안과학교실1, 서울대학교 의과대학 안과학교실2, 서울대학교 의과대학 분당서울대학교병원 안과학교실3
Abstract
PURPOSE
To investigate the accuracy of intraocular lens power calculations using simulated keratometry (simK) of dual Scheimpflug analyzer and 5 types of formulas in cataract patients. METHODS: The keratometry (K), axial length (AXL) and anterior chamber depth (ACD) were measured using ultrasound biometry (USB) combined with auto-keratometry (Auto-K), parital coherence interferometry (PCI; IOL master®) and dual Scheimpflug analyzer (DSA; Galilei®) in 39 eyes of 39 patients. Predicted refraction was calculated using Auto-K, mean K of PCI, and simK and total corneal power (TCP) of DSA in the Sanders-Retzlaff-Kraff (SRK-T) formula. The SRK-II, SRK-T, Holladay II, Haigis, and Hoffer-Q formula were used to calculate predicted refraction with the simK of DSA and AXL of USB. Manifest refraction, mean numerical error (MNE) and mean absolute error were evaluated 1, 3 and 6 months after cataract surgery. RESULTS: TCP of DSA was lower compared with other keratometric values (p < 0.05). The MNE was not different among Auto-K, mean K and simK. The MNE using TCP was larger compared with Auto-K, mean K and simK at 1 month after surgery (p < 0.05). There was a difference in MNE between simK and TCP of DSA at 6 months after surgery (p < 0.05). The MNE of SRK-T formula was the smallest in the intraocular lens (IOL) power calculation using the simK of DSA. CONCLUSIONS: We suggest using IOL power calculations with simK of DSA and SRK-T formula rather than TCP of DSA in cataract patients with normal corneas.
Key Words: Cataract surgery;Dual Scheimpflug analyger;Intraocular lens power calculation


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
#1001, Jeokseon Hyundai BD
130, Sajik-ro, Jongno-gu, Seoul 03170, Korea
Tel: +82-2-2271-6603    Fax: +82-2-2277-5194    E-mail: kos@ijpnc.com                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next