Measurement Comparison of Anterior Segment Parameters between AL-Scan(R) and Pentacam(R).
Sang Moon Youn, Sung Hyup Lim, Ho Young Lee
1Department of Ophthalmology, Inje University Busan Paik Hospital, Inje University College of Medicine, Busan, Korea.
2Shinsegae Eye Clinic, Busan, Korea. happytriad@gmail.com
에이엘 스캔과 펜타캠을 이용한 전안부 계측값의 비교
윤상문1⋅임성협1⋅이호영2
Department of Ophthalmology, Inje University Busan Paik Hospital, Inje University College of Medicine1, Busan, Korea
Shinsegae Eye Clinic2, Busan, Korea
Abstract
PURPOSE
To investigate the clinical availability of AL-Scan(R) (Nidek, GAMAGORI, Japan) by comparing anterior segment parameters measured with AL-Scan(R) and Pentacam(R) (Oculus, Wetzlar, Germany). METHODS: Seventy-three patients (117 eyes) who received refractive surgery at our hospital were tested with AL-Scan(R) and Pentacam(R). We compared measurements including anterior chamber depth, central corneal thickness, white-to-white, and corneal curvature. RESULTS: When comparing measurements obtained with AL-Scan(R) and Pentacam(R), the anterior chamber depth (p < 0.001), central corneal thickness (p < 0.001) and 2.4 mm zone K value (p = 0.038) showed significant differences; the white-to-white (p = 0.348) and 3.3 mm zone K value (p = 0.429) showed no significant differences. All AL-Scan(R) and Pentacam(R) parameters had a strong positive linear correlation (p < 0.001). The Bland-Altman plots showed a high degree of agreement between AL-Scan(R) and Pentacam(R) in all parameters except for anterior chamber depth. CONCLUSIONS: AL-Scan(R) is convenient to use clinically because simultaneous measurements of ocular biometry including axial length, intraocular lens power, and topography are possible. However, because differences in some anterior segment parameters exist when compared with Pentacam(R), measurements with AL-Scan(R) may require comparisons with other instruments.
Key Words: AL-Scan(R);Anterior chamber depth;Central corneal thickness;Corneal refractive power;Pentacam(R)


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next