Reliability and Usefulness of Refractive Measurements by PlusoptiX S09 in Children.
Jong Hoon Lim, Gi Hyun Bae, Seong Joo Shin
Department of Ophthalmology, Sahmyook Medical Center, Seoul, Korea. ymedi@hanmail.net
소아에서 플러스옵틱스 에스09로 측정한 굴절력의 신뢰성과 유용성
임종훈⋅배지현⋅신승주
Department of Ophthalmology, Sahmyook Medical Center, Seoul, Korea
Abstract
PURPOSE
To compare the refractive measurements obtained using a photorefractor (PlusoptiX S09, PlusoptiX GmbH, Germany) with those obtained using cycloplegic refraction in children. METHODS: We assessed the refractive status of 268 eyes in 134 children. The values acquired via photorefraction with a PlusoptiX S09 device were compared with those obtained by cycloplegic retinoscopy. Hyperopia (> or =+3.5 D), myopia (> or =-3.0 D), with the rule or against the rule astigmatism (> or =-1.5 D), and oblique astigmatism (> or =-1.0 D) were set as diagnostic criteria for refractive amblyopia risk factors (RARFs). The difference in the detection of RARFs by the two methods was the main outcome measure. RESULTS: The average spherical refractive power was -0.81 +/- 1.68 D for PlusoptiX S09 versus -0.26 +/- 2.00 D for cycloplegic retinoscopy (average difference -0.54 +/- 0.61 D; p < 0.001). The average spherical equivalent was -1.20 +/- 1.62 D for PlusoptiX S09 versus -0.64 +/- 1.94 D for cycloplegic retinoscopy (average difference -0.56 +/- 0.62 D; p < 0.001). The average cylinder power was -0.79 +/- 0.93 D for PlusoptiX S09 versus -0.76 +/- 0.94 D for cycloplegic retinoscopy (average difference -0.03 +/- 0.33 D; p = 0.135). Even though cycloplegic retinoscopy is considered the gold standard, the sensitivity and specificity for detecting RARFs with the PlusoptiX S09 were 88.0% and 96.3%, respectively. CONCLUSIONS: PlusoptiX S09 is a relatively useful method for detecting RARFs, but the device tends toward myopic shift compared to cycloplegic refraction, and hyperopia is underestimated.
Key Words: Amblyopia;Cycloplegic refraction;Photorefraction;PlusoptiX S09;Refractive amblyopia risk factors


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next