Comparison of Ocular Biometry and Postoperative Refraction in Cataract Patients between Galilei-G6(R) and IOL Master(R).
Jung Wook Lee, Seung Hun Park, Min Cheol Seong, Hee Yoon Cho, Min Ho Kang
Department of Ophthalmology, Hanyang University Guri Hospital, Hanyang University College of Medicine, Guri, Korea. ocularimmunity@gmail.com
갈릴레이 G6와 IOL 마스터를 이용한 안구생체계측과 백내장 술후 굴절력의 비교연구
이정욱⋅박승훈⋅성민철⋅조희윤⋅강민호
한양대학교 의과대학 한양대학교구리병원 안과학교실
Abstract
PURPOSE
To compare the axial lengths, anterior chamber depths, and keratometric measurements and to predict postoperative refractions of Dual Scheimpflug analyzer Galilei G6(R) and intra ocular lens (IOL) Master(R). METHODS: A total of 50 eyes in 50 patients who received cataract surgery were included in the present study. The axial length, anterior chamber depth, and keratometry were measured using 2 types of partial coherence interferometries (Galilei G6(R) and IOL Master(R)). The SRK/T formula was used to calculate IOL power and the predictive error which subtracts predictive refraction from postoperative refraction was compared between the ocular biometry devices. RESULTS: Axial lengths were 23.36 +/- 0.80 mm and 23.36 +/- 0.90 mm measured by Galilei G6(R) and IOL Master(R), respectively. Axial length measured by Galilei G6(R) was not statistically significant compared with IOL Master(R) (p = 0.321). The anterior chamber depth and keratometry were 3.22 +/- 0.35 mm and 44.29 +/- 1.40 D measured by Galilei G6(R) and 3.11 +/- 0.46 mm and 44.39 +/- 1.41 D measured by IOL Master(R), respectively. The differences of anterior chamber depth and keratometry between the 2 devices were statistically significant (p < 0.001 and p = 0.028, respectively). The mean absolute prediction errors were 0.45 +/- 0.37 D and 0.49 +/- 0.39 D in Galilei G6(R) and IOL Master(R), respectively and was not statistically significantly different (p = 0.423). CONCLUSIONS: The ocular biometric measurements and prediction of postoperative refraction using Galilei G6(R) were as accurate as with IOL Master(R).
Key Words: Axial length;Galilei G6(R);IOL Master(R);Partial coherence interferometry


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next