J Korean Ophthalmol Soc > Volume 54(9); 2013 > Article
Journal of the Korean Ophthalmological Society 2013;54(9):1345-1352.
DOI: https://doi.org/10.3341/jkos.2013.54.9.1345    Published online September 15, 2013.
Comparison of the Refractive Outcomes According to the Differences of Biometry and Keratometry Reading.
Kang Hoon Lee, Na Rae Kim, Kyoung Yul Seo
1Department of Ophthalmology and Inha Vision Science Laboratory, Inha University School of Medicine, Incheon, Korea.
2The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea. seoky@yuhs.ac
생체계측치와 각막 곡률의 차이에 따른 인공수정체 삽입 후 굴절 오차의 경향
이강훈1⋅김나래1⋅서경률2
Department of Ophthalmology and Inha Vision Science Laboratory, Inha University School of Medicine1, Incheon, Korea The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine2, Seoul, Korea
Abstract
PURPOSE
The purpose of this study was to investigate the error tendency between preoperative expected refraction and postoperative manifest refraction based on axial length, anterior chamber depth, and keratometric data obtained by an automated keratometer and Pentacam(R) in cataract surgery cases and to report how their differences affect determination of intraocular lens (IOL) power. METHODS: The authors retrospectively reviewed the medical records of 110 eyes of 84 patients who underwent cataract surgery. Axial length and anterior chamber depth were measured by A scan ultrasound biometry, while keratometric values were obtained by an automated keratometer and Pentacam(R). IOL power was calculated using the SRK/T formula. Patients were divided into 3 groups based on the axial length, anterior chamber depth, and the difference of keratometric values between the 2 devices. Refractive error was analyzed 2 months after surgery. RESULTS: There were no statistically significant differences between axial length and anterior chamber depth among the groups; however, the K reading differences were statistically significant. Although the mean absolute error (MAE) of each group showed no statistical significance among the groups, the MAE was more pronounced in the group in which the keratometeric value measured by Pentacam(R) differed more than 1.00 diopter from the automated keratometer measurements. CONCLUSIONS: There was no statistically significant difference between axial length and anterior chamber depth among the groups. A difference of 1.00 diopter or more between the keratometric values obtained by an automated keratometer and Pentacam(R) significantly affects the postoperative refractive error; therefore, these factors should be considered when determining IOL power.
Key Words: Biometry;Cataract surgery;Keratometry reading;Refractive outcomes;Scheimpflug photography


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next