Axial Length Change after Implantable Collamer Lens Implantation.
Ju Yong Seok, Damho Lee, Haksu Kyung, Joon Mo Kim
1Department of Ophthalmology, National Medical Center, Seoul, Korea.
2Vision Eye Center, Seoul, Korea. damholee@naver.com
3Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Korea.
ICL 삽입술 후 안축장 측정값의 변화
석주용1⋅이담호2⋅경학수1⋅김준모3
Department of Ophthalmology, National Medical Center1, Seoul, Korea
Vision Eye Center2, Seoul, Korea
Department of Ophthalmology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine3, Seoul, Korea
Abstract
PURPOSE
The axial length measurement is an important variable in calculating the intraocular lens (IOL) power in cataract surgery. In this study, the effects of the implantable collamer lens (ICL) on the axial length measurement were investigated. METHODS: The axial lengths of 141 eyes of 76 highly myopic patients were measured before ICL implantation and 1 month postoperatively using applanation A-scan ultrasound. RESULTS: The measured axial lengths before ICL implantation and 1 month after the surgery were 28.03 +/- 1.91 mm and 28.08 +/- 1.86 mm respectively, with a statistically significant difference of 0.05 mm (p = 0.036). These 2 values showed a high correlation (R2 = 0.9829, p < 0.0001). CONCLUSIONS: Even though the axial length after ICL implantation significantly increased, the 0.05 mm change was too small to influence the determination of the IOL diopters in cataract surgery. Therefore, ICL implantation history may be reflected when measuring the axial length for the IOL power calculation in cataract surgery, but the axial length measured immediately before the cataract surgery may be used to calculate the IOL power regardless of the ICL implantation history.
Key Words: A-scan ultrasonography;Axial length measurement;Implantable collamer lens


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next