Comparison of Optical Quality Between Two Intraocular Lenses Using Double-Pass Based Optical Quality Analysis System.
Ae Young Kwak, Moonjung Choi, Kyoung Yul Seo, Ji Min Ahn
1The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.
2Siloam Eye Hospital, Seoul, Korea. 24mini@naver.com
두 종류 비구면 인공수정체 삽입 후 Double-pass 원리를 이용한 시력의 질 비교
곽애영1,2⋅최문정1⋅서경률1⋅안지민2
The Institute of Vision Research, Department of Ophthalmology, Yonsei University College of Medicine1, Seoul, Korea
Siloam Eye Hospital2, Seoul, Korea
Abstract
PURPOSE
To compare postoperative optical qualities between two types of 1-piece aspheric intraocular lenses using the double-pass technique. METHODS: Uncorrected visual acuity (UCVA), best corrected visual acuity (BCVA), and spherical equivalent were evaluated in the subject groups; the first which included 25 eyes implanted with Tecnis(R) ZCB00 and the second which included 16 eyes implanted with Acrysof(R) IQ SN60WF. In addition, modulation transfer function (MTF) cut-off, Strehl ratio, and objective scattering index (OSI) were measured 6 months after cataract surgery in the 2 subject groups using Optical Quality Analysis System (OQAS, Visiometrics S.L., Terrasa, Barcelona, Spain) which is based on the double-pass technique. RESULTS: There were no significant differences in the spherical equivalent, UCVA, BCVA, and OSI between the 2 groups. However, both the MTF cut-off, and Strehl ratio showed statistically significant differences. The MTF cut-off (28.0 +/- 7.79 vs. 20.4 +/- 9.51 c/deg, p = 0.025) and Strehl ratio (0.14 +/- 0.04 vs. 0.12 +/- 0.05, p = 0.042) were higher in the Tecnis(R) ZCB00-implanted group. CONCLUSIONS: The difference in characteristics of intraocular lenses subtly affects the vision quality as measured by values such as MTF cut-off and Strehl ratio after cataract surgery. OQAS based on the double-pass technique is considered useful in more objective estimates of the real retinal image quality after cataract surgery which is difficult to explain simply by measuring visual acuity.
Key Words: Double-pass;Optical Quality Analysis System;Visual quality;1-piece aspheric intraocular lenses


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next