J Korean Ophthalmol Soc > Volume 52(5); 2011 > Article
Journal of the Korean Ophthalmological Society 2011;52(5):610-617.
DOI: https://doi.org/10.3341/jkos.2011.52.5.610    Published online May 15, 2011.
The Measurement of Flow Resistance in Drainage Implants Using Various Tube Ligation Methods.
Hyung Ju Park, Ki Ho Park, Seok Hwan Kim, Tae Woo Kim, Dong Myoung Kim
Department of Ophthalmology, Seoul National University College of Medicine, Seoul, Korea. kihopark@snu.ac.kr
튜브 결찰 및 스텐트 방식에 따른 방수유출장치 내 유동저항의 평가
박형주⋅박기호⋅김석환⋅김태우⋅김동명
Department of Ophthalmology, Seoul National University College of Medicine, Seoul, Korea
Abstract
PURPOSE
To evaluate the flow resistance though the tube in a glaucoma drainage implant using various tube ligation methods. METHODS: To measure the flow resistance, the following tube ligation methods were designed and tested: Type 1: no ligation of Ahmed valve tube. Type 2: ligation of Ahmed valve tube with 8-0 Vicryl. Type 3: ligation of Molteno tube and a 6-0 Vicryl as an intra-luminal stent with 8-0 Vicryl. Type 4: ligation of Ahmed valve tube and three strands of 8-0 nylon as extra-luminal stents with 8-0 Vicryl. Type 5: ligation of Ahmed valve tube and a 6-0 Prolene as an extra-luminal stent with 8-0 Vicryl followed by removal of the Prolene for partial ligation. RESULTS: The pressure was maintained under 0.143 mm Hg in type 1. In type 2, the pressure increased to 6.688 mm Hg and dropped to approximately 6.384 mm Hg. In type 3, the pressure was maintained at 5.396 mm Hg which decreased to 3.572 mm Hg after stent removal. In type 4, the pressure was maintained at 5.700 mm Hg which dropped to 5.472 mm Hg after the 1st stent removal, to 5.016 mm Hg after the 2nd stent removal and to 4.180 mm Hg after the 3rd stent removal. In type 5, the pressure increased to 6.384 mm Hg and decreased to 5.624 mm Hg. CONCLUSIONS: The tube ligation along with extra-luminal stents followed by staged removal may provide prevention of hypotony and staged control of intraocular pressure after a glaucoma implant operation.
Key Words: Ahmed valve;Ligation;Molteno implant;Stent


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next