Comparison of Aberrations in Korean Normal Eyes Measured With Two Different Aberrometers.
Jung Hoon Yum, Suk Kyue Choi, Jin Hyoung Kim, Do Hyung Lee
Department of Ophthalmology, Ilsan Paik Hospital, Inje University College of Medicine, Goyang, Korea. eyedr0823@hotmail.com
한국인 정상안을 대상으로 서로 다른 수차계로 측정된 안구 수차 값의 비교
염정훈ㆍ최석규ㆍ김진형ㆍ이도형
Department of Ophthalmology, Ilsan Paik Hospital, Inje University College of Medicine, Goyang, Korea
Abstract
PURPOSE
To compare ocular higher order aberrations measured by two different aberrometers in a sample of normal eyes. METHODS: We included 41 normal eyes of Koreans in this study. Ocular aberration data were obtained through three measurements per eye using Zywave and OPD-Scan devices. Spherical equivalent and higher order aberrations calculated in the central 6 mm zone and expressed as root mean square (RMS) values were analyzed. RESULTS: A comparison of measurements between the Zywave and OPD-Scan devices demonstrated no statistically significant differences in the RMS values of total higher order aberration (p=0.11), but significant differences were detected in the RMS values of total spherical aberration, total coma and total trefoil (p<0.01). CONCLUSIONS: The two different aberrometers that we tested are suitable for taking repeated measurements and are internally consistent, but not interchangeable.
Key Words: Aberrometer;Higher order aberration;Normal eye


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next