J Korean Ophthalmol Soc > Volume 51(6); 2010 > Article
Journal of the Korean Ophthalmological Society 2010;51(6):829-834.
DOI: https://doi.org/10.3341/jkos.2010.51.6.829    Published online June 15, 2010.
Stability of Four-Haptic Intraocular Lens in Combined Phacoemulsification and Vitrectomy.
Jae Woo Kim, Ji Wook Yang, Dong Hyun Jee
Department of Ophthalmology and Visual Science, The Catholic University of Korea School of Medicine, Suwon, Korea. Dhjee73@nate.com
유리체절제술과 백내장 동시수술에서 네지지부인공수정체의 안정성
김재우ㆍ양지욱ㆍ지동현
Department of Ophthalmology and Visual Science, The Catholic University of Korea School of Medicine, Suwon, Korea
Abstract
PURPOSE
To evaluate the stability of intraocular lenses (IOLs) according to the haptic design in combined surgery (phacoemulsification, intraocular lens insertion, and vitrectomy) by comparing early IOP elevation and the accuracy of IOL power. METHODS: A retrospective study was conducted for 85 eyes of 85 patients who had undergone combined surgery. IOP elevation one week postoperatively and the difference between target refraction and postoperative refraction after two months were evaluated in all patients. All eyes were allocated to either a 2-haptic IOL group or a 4-haptic IOL group, and then each group was divided into three subgroups: gas-filled, silicone oil-filled, or balanced salt solution-filled. RESULTS: In the balanced salt solution-filled group, the early IOP increase was significantly less in the 4-haptic IOL group than it was in the 2-haptic IOL group (3.24 mmHg vs. 6.15 mm Hg, p=0.025). The myopic shifts of the 2-haptic IOL group were significantly greater than those of the 4-haptic IOL group for all of the subgroups (gas subgroup -0.71D vs. -0.31D, p=0.045; balanced salt solution subgroup -0.66D vs. -0.16D, p=0.018). CONCLUSIONS: In combined surgery, 4-haptic IOL implantation showed significantly less early IOP elevation and less myopic shift than those of 2-haptic IOL implantation.
Key Words: Combined surgery;4-haptic IOL;Intraocular pressure;Myopic shift


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next