Comparison of Phacoemulsification Time: Oscillation Device versus Conventional Device.
Min Kim, Seung Hee Lee, Hae Young Lee, Sung Kun Chung
1Department of Ophthalmology Seoul Adventist Hospital, Seoul, Korea.
2Department of Ophthalmology and Visual Science, St. Mary's Hospital, The Catholic University of Korea, Seoul, Korea. eyedoc@catholic.ac.kr
회전성 진동 손잡이를 이용한 수정체유화술의 수정체유화시간에 대한 비교
김민1,이승희1,이해영1,정성근2
Department of Ophthalmology Seoul Adventist Hospital1, Seoul, Korea, Department of Ophthalmology and Visual Science, St. Mary`s Hospital2, The Catholic University of Korea, Seoul, Korea
Correspondence:  Sung Kun Chung, M.D.
Abstract
PURPOSE
To compare the phacoemulsification time used in an adjunctive oscillation device and a conventional ultrasound phacomachine. METHODS: Thirty eyes had phacoemulsification with a NeoSoniX(R) (Alcon Infiniti phacoemulsification system, Alcon, TX, U.S.A.) tip, and 30 eyes had phacoemulsification with a conventional ultrasound tip. Using the Scheimpflug camera (EAS-1000, Nidek, Japan), densities of the lens were measured in the eye of each patient in both groups preoperatively. The setting of parameters for phacoemulsification in both groups was identical. The power was 19% and the vacuum setting was 180 mmHg. A pulse rate of 40 pulses per second was used. For the cataract surgery with NeoSoniX(R), the setting of the oscillation angle was fixed to 100% (4 degrees). After cataract extraction, the phacoemulsification time was recorded. RESULTS: The mean phacoemulsification time was 28.50+/-12.40 seconds in the NeoSoniX(R) group and 49.87+/-45.55 seconds in the conventional group. The reduced phacoemulsification time was statistically significant (p<0.05) in the NeoSoniX(R) group. CONCLUSIONS: There was a statistically significant difference of phacoemulsification time between the NeoSoniX(R) and conventional phacoemulsification groups. Therefore, NeoSoniX(R) may contribute to a reduction in phacoemulsification time in all applications and can be an effective adjunctive device in a conventional phacomachine.
Key Words: NeoSoniX(R);Oscillation device;Phacoemulsification time


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next