Pachymetric Measurements Using Orbscan after Excimer Refractive Surgery.
Minjae Kwon, Youlseok Seong, Dalwoong Huh
Department of Ophthalmology, Kangbuk Samsung Hospital, College of Medicine, Sungkyunkwan University, #108 Pyung-dong, Jongro-gu, Seoul, Korea. huhdalwoong@hanmail.net
엑시머 레이저를 이용한 각막 굴절교정수술 후 Orbscan 을 이용한 각막두께 측정
권민재 ( Min Jae Kwon ) , 성열석 ( Youl Seok Seong ) , 허달웅 ( Dal Woong Huh )
Abstract
PURPOSE
To assess the accuracy of pachymetric measurement using Orbscan(Bausch and Lomb, USA) after laser in situ keratomileusis(LASIK), photorefractive keratectomy(PRK), and laser epithelial keratomileusis (LASEK). METHODS: Central corneal thickness of 159 nonoperated normal eyes, 55 eyes after LASIK, 12 after PRK, and 17 after LASEK was measured using ultrasound and Orbscan II. The acoustic factor (AF) was adjusted, based on the results obtained in the normal eye group, to minimize the difference between ultrasound and Orbscan pachymetric values. RESULTS: Using the adjusted AF (0.9607), the mean difference between Ultrasonic pachymetric and Orbscan measurement was 0.00 +/- 21.08, 20.44 +/- 24.12, 27.63 +/- 36.87, and 0.56 +/- 17.26 micro meter in the normal, LASIK, PRK, and LASEK groups, respectively. In the LASIK and PRK groups, there was a statistically significant difference between the two methods (both P<0.001). The postoperative ultrasound measurements were consistent with the theoretical residual corneal thickness in the LASIK and PRK groups (-2.96 +/- 18.00 micro meter, -1.00 +/- 13.59 micro meter), while Orbscan measurements were statistically significantly less than the theoretical residual corneal thickness in the LASIK and PRK groups (21.84 +/- 31.06 micro meter, p<0.001; 21.88 +/- 35.91 micro meter, P=0.036). CONCLUSIONS: Orbscan pachymetric values may be underestimated and less accurate after excimer refractive surgery, especially after LASIK and PRK.
Key Words: Acoustic factor;AF;Corneal haze;Corneal thickness;Excimer refractive surgery;LASIK;LASEK;Orbscan;Pachymetry;PRK


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next