Corneal Topographic Evaluation of Centration of Excimer Laser Myopic Photorefractive Keratectomy.
Sang Bumm Lee, Hee Kyu Kim
Department of Ophthalmology, College of Medicine, Yeungnam University, Taegu, Korea.
엑시머레이저 근시교정술시 연마중심 ( Ablation Center ) 에 대한 각막형태검사를 이용한 평가
이상범(Sang Bumm Lee),김희규(Hee Kyu Kim)
Abstract
In order to get a good visual outcome following excimer laser photorefractive keratectomy(PRK) the centration of ablation is very important. In this study, the ablation center of excimer laser PRK using VISX 20/20 B(R) with VisionKeyTM(VISX, Inc. Santa Clara, CA) was evaluated by using corneal topography. Corneal topography was performed before operation and one month after operation to 190 eyes of 123 patients. The location and distance of center of ablation from center of the entrance pupil were calculated by using EyeSys Corneal Analysis System(R) with pupil-finding software(EyeSys Laboratories, Inc. Houston, TX). Supero-nasal displacement of ablation center occurred in 84 eyes(44%) after excimer laser PRK. The mean distance was 0.33 +/- 0.21mm(range: 0.02~1.36mm). The distance less than 0.5mm was found in 157 eyes(83%) and that between 0.5mm and 0.75mm was found in 26 eyes(14%). The decentering amount following excimer laser PRK was not depending on laterality(OD vs. OS), sex, age, amount of attempted correction(6D down vs. 6D up), type of software(ver 3.20 vs. 4.01) or type of procedure(PRK vs. PARK). In contrast, the second eye receiving excimer laser PRK showed a better centration of ablation than the first eye. And the centration results improved with surgeon's experience in the most recent 50 eyes(0.28 +/- 0.16mm). However, there was no relation between amount of decentration and visual improvement. In this study, it was found that decentering of ablation center less than 0.75mm did not influence the visual outcome in excimer laser PRK.
Key Words: Centration/Decentration;Cornea;Corneal topography;Entrance pupil;Excimer laser PRK


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next