Measurement of Corneal Power and Astigmatism Using Placido-based Videokeratography and Comparison with Other Keratometers.
Yung Hui Kim, Ying Li, Hyo Seok Lee, Kyung Chul Yoon
Department of Ophthalmology, Chonnam National University Medical School, Gwangju, Korea. kcyoon@jnu.ac.kr
플라시도원반 방식의 비디오각막경을 이용한 각막곡률 및 난시측정과 기존의 각막곡률계와의 비교
김영휘⋅이 영⋅이효석⋅윤경철
전남대학 의과대학 안과학교실
Abstract
PURPOSE
In the present study, the repeatability and reproducibility of the corneal power and astigmatism measurements using placido-based video keratography were evaluated and the agreement with other keratometers were compared. METHODS: This prospective study included 45 patients (45 eyes) scheduled to undergo cataract surgery between November 2015 and January 2016. Three sets of corneal power and astigmatism were measured using placido-based video keratometer (Keratograph® 5M), automatic keratometer (KR-8900®), manual keratometer (B×L manual keratometer®), Placido-scanning-slit keratometer (ORBscan II®), Scheimpflug keratometer (Pentacam®), and low coherence interferometry (Lenstar LS900®). Reliability of each device was analyzed using the coefficient of variation, standard deviation and intraclass correlation coefficient. Repeated measures analysis of variance was used to analyze the interdevice comparison of mean absolute difference. The agreement between the devices was evaluated with 95% limits of agreement (LoA) and Bland-Altman plots. RESULTS: The mean corneal power, astigmatism and power vector analysis (J₀, J₄₅) were not significantly different among devices (p > 0.05). In the Bland-Altman plot analysis, the 95% LoA of corneal power, J₀, and J45 when comparing Keratograph® 5M with others ranged from -0.78 to 0.55 D, from -0.42 to 0.45 D, and from -0.39 to 0.47 D, respectively. CONCLUSIONS: Keratograph® 5M showed good repeatability and reproducibility of corneal power and astigmatism measurements and was interchangeable with other keratometers.
Key Words: Astigmatism;Corneal power;Keratograph;Keratometer;Placido-based


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next