Comparison of Intraocular Pressures Measured by the Corvis ST and Other Tonometers in Normal Eyes.
Seung hoon Lee, Jung il Moon, Youn Hea Jung
Department of Ophthalmology, College of Medicine, The Catholic University, Seoul, Korea. younhea@hotmail.com
정상 한국인에서 Corvis ST와 골드만압평안압계, 리바운드안압계로 측정한 안압 비교
이승훈 · 문정일 · 정윤혜
가톨릭대학교 의과대학 안과학교실
Correspondence:  Youn Hea Jung,
Email: younhea@hotmail.com
Received: 22 May 2019   • Revised: 10 June 2019   • Accepted: 6 December 2019
Abstract
PURPOSE
We compared the intraocular pressures (IOPs) measured by the Corvis ST (CST), Goldmann applanation tonometer (GAT), and Rebound tonometer (RBT) and correlated the measured IOPs with the corneal center thickness (CCT). METHODS: The IOPs were measured in 34 eyes of 34 normal subjects using the CST, GAT, and RBT by independent examiners. Comparisons between the IOPs measured by the CST and other tonometers were made using paired t-tests and Bland-Altman plots and Pearson's correlation coefficient was used to correlate the measured IOPs with the CCTs. RESULTS: The IOP measured by the CST (C-IOP; 14.57 ± 2.37 mmHg) was significantly higher than that measured by the GAT(13.56 ± 2.25 mmHg) and RBT (13.57 ± 3.26 mmHg) (p < 0.05). The biomechanically corrected C-IOP (C-bIOP; 13.40 ± 2.06mmHg) showed no difference from the GAT-IOP and RBT-IOP (p > 0.50). In analyzing the correlation with CCT, RBT-IOP (R =0.34; p = 0.705) showed a lower correlation than the GAT-IOP (R = 0.49; p < 0.01) and C-IOP (R = 0.48; p < 0.01). The C-bIOP showed no correlation with the CCT (p = 0.74). CONCLUSIONS: The IOP measurements using the C-IOP were significantly higher than the RBT-IOP and GAT-IOP but the C-bIOP showed no difference from the RBT-IOP and GAT-IOP. C-IOP, RBT-IOP, and GAT-IOP were significantly correlated with corneal thickness, but there was no correlation between CCT and C-bIOP. C-bIOP may therefore be considered clinically useful.
Key Words: Biomechanical corrected intraocular pressure;Central corneal thickness;Corvis ST


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
SKY 1004 Building #701
50-1 Jungnim-ro, Jung-gu, Seoul 04508, Korea
Tel: +82-2-583-6520    Fax: +82-2-583-6521    E-mail: kos08@ophthalmology.org                

Copyright © 2024 by Korean Ophthalmological Society.

Developed in M2PI

Close layer
prev next